Functions | Time Interval (between pulses at up to eight inputs or pulses appearing consecutively at a single, common input) Time Interval Error, Maximum Time Interval Error, Time Deviation |
Frequency | Allan Deviation |
Statistics | Mean, Min and Max Values, Standard Deviation |
Time Interval | Range 73 min Resolution (LSB) 1.9 ps in single-shot measurements, Independent, Common High Speed and Source Precision (Standard Deviation) < 6.5 ps at time interval measured from 0 to 500 ms (internal rubidium) up to 4.5 ps in Common High Precision mode |
Frequency & Period | Range Inputs 1 - 8: 1 mHz to 250 MHz Sensitivity < 75 mV RMS typ. (0.01 to 250 MHz) Minimum slew rate: 10 V/μs Input F: 100 MHz to 3.5 GHz Sensitivity < -12 dBm (< 55 mV RMS) from 400 MHz to 3 GHz Sensitivity < - 3 dBm (< 160 mV RMS) from 100 MHz to 3.5 GHz Gate Time selected from 1 μs to 10 s (reciprocal method) Measurement Rate up to 106 measurements/sec (depends on selected gate) |
Inputs 1 - 8 | Impedance: 50 W, DC coupled; SMA sockets Amplitude: within ± 4 V Pulse edge: selectable, rising or falling Threshold: manually adjustable from -4 V to +4 V with8 mV resolution, or set automaticall |